Sensorwell

7541-112

These advanced instruments provide engineering unit display of a strain gage (mV/V) input and a frequency input. They also compute power and perform 21 functions including limit checks, tare, hold and max./min. capture. You needn’t write code or add hardware to be up and running a productive test. The alphanumeric readout can display measured and computed data, units of measure, and test status. During setup, it guides you with English language prompts. There are no manual adjustments. To calibrate, enter the full scale value in engineering units and auto-cal takes over. Provides 0.01 % (max.) resolution and ±5 V and/or ±10 V analog outputs at full scale. The keyboard accesses measured data, held data, max./min. data, data spread, limit status, and/or I/O status without disruption. Password protection may be used, if needed.

Request now

7541-112

Produkt Merkmale

Kenndaten

  • May be calibrated in two directions (clockwise and counter-clockwise or tension and compression)
  • Fast, rock-solid readings with high noise immunity – 2000 samples/sec for torque, head, or drawbar force input – 1 ms response for speed
  • One or two channel input
  • Six-digit engineering unit with legends and 0.01 % resolution max.
  • RS-232, RS-422, RS-485 serial communication
  • Auto-scaled ±5 V and/or ±10 V analog outputs
  • No pots, batteries, fans, maintenance, or external power supplies
  • Excitation: ac carrier models, 3.28 kHz, 3 V rms; dc models, 5 V or 10 V dc, user selectable
  • CE approved

Technische Daten

Order Code
7541-112
Input Type
Any 80 Ohm to 2K Ohm transducer, directly wired or transformer coupled. 4, 6, or 7 wire circuits are accommodated
Display Type
LCD
Numeric Display Format
16 x 2
Transducer Excitation
ac excitation: 3 V RMS, 3.28 kHz ±0.01 % sine wave. Regulated and short circuit protected
Type of Inputs
Dual display (two inputs)
Sensitivity
1.0 mV/V to 4.5 mV/V with 50 % overrange; automatically scaled
Supply Voltage
90 Vac to 250 Vac
Input Resistance
100 mohm in parallel with 33 pF
Automatic Null
In Phase: ±10 % of full scale (with 50 % overrange), ±60 % of full scale (with 0 % overrange), Quadrature: ±1 mV/V
Auto Calibration
Dual polarity shunt calibration with provision for CAL resistor feedback
Spurious Signal Rejection
60 Hz: 120 dB common mode, 100 dB normal mode. Carrier quadrature: 60 dB
Anti-Alias Filter
200 Hz, 7-pole Bessel response filter
Low-Pass Filtering
20 kHz (-3 dB) or none (this filter is not available for TTL inputs)
Resolution
0.01 % of full scale
Overall Accuracy
0.01 % of full scale @ 25 °C [77 °F]; 0.015 % of full scale @ 5 °C to 50 °C [41 °F to 122 °F]
Frequency Input
Any unidirectional or bidirectional (quadrature) source including self-generating and zero velocity magnetic pickups, optical encoders, flowmeters, etc. When used with bidirectional sensors, the conditioner outputs both direction and magnitude
Input Impedance and Configuration
Differential or single-ended inputs. 100 kohm differential, 50 kohm single ended
Input Threshold (Keypad Selectable)
10, 20, 50, 100, or 200 mV pk-pk (between inputs) or TTL
Maximum Voltage
±130 Vdc or 130 Vrms
Input Signal Bandwidth
0.001 kHz to 200 kHz (10 mV to 200 mV pk-pk threshold); 0.001 kHz to 400 kHz (TTL threshold)
Display Ranges and Resolution
Rangeless (use any full scale engineering unit value) with 50 % overrange. Resolution is 0.01 % of full scale
Maximum Transducer Cable Length
500 ft, except 200 ft for 100 ohm or lower strain gage transducers
Signal-to-Noise Ratio
With 1/10/100/200 Hz filters 86/76/66/62 dB @ 1 mV/V full scale and 86/80/72/66 dB @ 5 mV/V FS
Maximum Baud Rate
300 to 38400. Maximum cable length: 4000 ft (RS-422/RS-485), 50 ft (RS-232)
Response Time
Greater of: 1 ms, typ. (2 ms worst case) or the input pulse length
Overrange Indicator
±8.2 V @ ±5 V full scale or ±13.5 V@ ±10 V full scale
Common Mode Rejection
80 dB (60 Hz), 55 dB (0 kHz to 10 kHz)
Weight
1,36 kg [3 lb]
Product Dimensions
165 mm W x 74 mm H x 221 mm D [6.5 in W x 2.9 in H x 8.7 in D]
UNSPSC Code
41111605
UNSPSC Commodity
Strain gages
Availability
Global
Series Name
7541

We are happy to help you!

Inquire now

Our partners

Kalsdorfer Straße 26 a
8073 Feldkirchen bei Graz
Austria

Tel:    +43 316 40 33 01
Mail:  office@sensorwell.at